A clamp that read half
The jaw was not closed. The afternoon log was half of the morning log. Nothing in the panel had changed.
Xiyun Instruments records what the meter showed, what the reference cell showed, and which reading we kept.
The jaw was not closed. The afternoon log was half of the morning log. Nothing in the panel had changed.
A two-degree offset showed up below 5°C. It disappeared in the office. The sensor stayed on the north wall.
The logger stored 10.26. The screen showed 10.3. The report used the screen until we checked the file.
No samples between 02:00 and 02:14. The clock battery was the cause. The meter itself was fine.
We log a full day before changing a setting. Serial numbers and firmware versions go in the notebook first. There is no portal and nothing here to sign in to.